Model: CCT-8302A
Measurement items: Double channels
Auxiliary electrode constant: 10.0、1.0、0.1 、 0.01(cm-1 )
Measurement range: Conductivity :0.05μS/cm ~150.0mS/cm
Resistivity : 0.02 MΩ·cm~18.25 MΩ·cm
TDS:0.02ppm~100ppt
Temperature :0~99.9℃
Resolution: Resistivity: 0.01 MΩ·cm
Conductivity:0.01μS/cm 0.1% of readings is equal to four digits,
TDS:0.01ppm
Temperature :0.1℃
Accuracy:
TDS
Resistivity
Conductivity :1.0%(FS)
Temperature:±0.5℃
Temperature compensation:
Temperature components PT1000, Digital calculation compensation
Electrode coefficient:
0.008~12.000cm-1
Current output:
double channels isolated 4-20ma output
( conductivity/temperature for selection)
Maximum loop resistance500Ω
(Accuracy: ±0.1mA or±1.0%of full scale)
Control output:
Triple channels isolated configuration OC output, external power
supply, Load current 50mA ,voltage ≤36V
Protection Level: IP65(with rear cover)
Working conditions: Temperature:0 ~ 50 ℃;Humidity:≤80% RH
Power supply: AC/DC 18~36V(ripple Vp-p <100mV)
Dimension: 96mm×96mm×95mm(H×W×D)
Slot dimension: 91mm×91mm(H×W)
Installation: panel mounted